  • Data for: Evaluation of the Beckman Coulter DxC 700 AU Chemistry Analyzer
    The supplemental data is related to the article entitled “Evaluation of the Beckman Coulter DxC 700 AU Chemistry Analyzer”. The data provides one lab’s performance with the new Beckman Coulter DxC AU 700 compared to an older Siemen’s Vista 500. Performance characteristics included intra and inter-run precision, linearity/analytical measurement range, method correlation, reference range verification and calculated sigma values. A total of 53 assays including 11 critical care, 19 general chemistries, 11 proteins, 10 urines, and 2 CSF analytes were tested. Herein is a summary of the data for all analytes tested.
    • Dataset
  • Data for: Effects of different tube types on patient classification using current diabetes decision limits
    stability testing of glucose and lactate in tubes with different glycolysis inhibitors from Greiner and Sarstedt. Study performed in volunteers. Storage of samples for up to 48 hrs, centrifugation either immediately after blood drawing or delayed for 48 hrs.
    • Dataset
  • Data for: Comparison of Analytical Performance of i-Smart 300 and pHOx Ultra for the Accurate Determination of Pleural Fluid pH
    We compared anerobically collected pleural fluid pH by iSTAT, pHOx Ultra, i-Smart blood gas analyzers and pH papers. Dataset are described in the attached excel file and analyzed sheet by Analyzed-it software.