Paired serum neurofilament light chain measurements on Siemens Atellica IM and Roche Elecsys platforms

Published: 12 October 2026| Version 1 | DOI: 10.17632/8bwvvh4yp2.1
Contributor:
DANIEL FATELA CANTILLO

Description

This dataset contains paired serum neurofilament light chain (sNfL) concentrations measured in the same 133 serum samples on two routine immunoassay platforms, the Siemens Atellica® IM Neurofilament Light Chain assay and the Roche Elecsys® NfL assay on cobas® e801, within a single analytical session. Each record comprises the concentration obtained on each platform in pg/mL and its base-10 logarithm, together with the age at sampling, sex and diagnostic group of the participant (0 = healthy control, n=28; 1 = multiple sclerosis, n=99; 2 = other neurological diagnosis, n=6). Patient samples were residual material remaining after routine clinical testing and healthy control samples were obtained with written informed consent; all records are pseudonymized and contain no direct identifiers. The dataset supports the method comparison, Passing-Bablok regression, Bland-Altman analysis and cross-platform classification analyses reported in the associated article.

Files

Categories

Clinical Biochemistry, Clinical Neurology