Data for: Validation of ABI 3500xL Genetic Analyzer after decommissioned and recommissioned at new permises

Published: 7 Oct 2017 | Version 1 | DOI: 10.17632/h3bh292rcg.1

Description of this data

This validation was aimed to demonstrate that there is no effect on the Applied Biosystems® 3500xL Genetic Analyzer (Thermo Fisher Scientific®, UK) after decommissioned and recommissioned at new premises. Same sample sets were tested on the ABI 3500xL Genetic Analyzer at both old and new premises using current accredited procedures. The results showed genotyping concordance which means using the ABI 3500xL Genetic Analyzer at the new premises provides no risk regarding the accuracy, precision and reproducibility of the results and also showed a very good performance on sensitivity, specificity, repeatability and good profile quality based on the assessment of locus peak balance and peak height.

Experiment data files

This data is associated with the following publication:

Validation of ABI 3500xL Genetic Analyzer after decommissioned and recommissioned at new premises

Published in: Forensic Science International: Genetics Supplement Series

Latest version

  • Version 1

    2017-10-07

    Published: 2017-10-07

    DOI: 10.17632/h3bh292rcg.1

    Cite this dataset

    Iyavoo, Sasitaran; Haizel, Thomas; Draus-Barini, Jolanta (2017), “Data for: Validation of ABI 3500xL Genetic Analyzer after decommissioned and recommissioned at new permises”, Mendeley Data, v1 http://dx.doi.org/10.17632/h3bh292rcg.1

Statistics

Views: 187
Downloads: 5

Categories

DNA, Capillary Electrophoresis, Accreditation, Validation Study

Licence

CC BY 4.0 Learn more

The files associated with this dataset are licensed under a Creative Commons Attribution 4.0 International licence.

What does this mean?

This dataset is licensed under a Creative Commons Attribution 4.0 International licence. What does this mean? You can share, copy and modify this dataset so long as you give appropriate credit, provide a link to the CC BY license, and indicate if changes were made, but you may not do so in a way that suggests the rights holder has endorsed you or your use of the dataset. Note that further permission may be required for any content within the dataset that is identified as belonging to a third party.

Report