High-resolution digital densitometry dataset for lubricant method verification under ISO/IEC 17025:2017
Description
This dataset contains 168 primary density measurements across seven hydrocarbon matrices (3 certified reference materials, 2 Group II base oils, and 2 formulated lubricants) obtained using ultra-high-resolution oscillating U-tube densitometry (Anton Paar DMA 4500 M, resolution: 10⁻⁵ g/cm³). The data supports the study "The parametric-nonparametric paradox in high-resolution analytical data: A robust statistical framework for precision evaluation and decision-making in digital densitometry" submitted to Chemometrics and Intelligent Laboratory Systems. The dataset includes raw measurements, descriptive statistics, normality tests (Shapiro-Wilk), homogeneity tests (Levene, Cochran), parametric (ANOVA) and non-parametric (Kruskal-Wallis) hypothesis testing results, precision parameters, bias assessment, GUM uncertainty budgets, and uncertainty propagation to derived parameters (relative density and API gravity).
Files
Institutions
- Escuela Superior Politecnica del LitoralGuayas, Guayaquil